BIOPAC Epoch Wireless EEG Transmitters

Dual-channel implantable EEG/Neural wireless transmitter for pup, mouse, or rat. The wireless transmitter amplifies and transmits two channels of high-fidelity EEG data to an EPOCH System receiver tray for long-term wireless EEG recording from small animals in their home cages.

BIOPAC Epoch Wireless EEG Transmitters

Part #: EPOCH-T2, EPOCH-T6, EPOCH-P-EEG-SEN

Dual-channel implantable EEG/Neural wireless transmitter for pup, mouse, or rat. The wireless transmitter amplifies and transmits two channels of high-fidelity EEG data to an EPOCH System receiver tray for long-term wireless EEG recording from small animals in their home cages.

·         Small implants uniquely suited for neonatal and adult rodent recording

·         Two-week transmitter suitable for pre-weaned rodents (less than postnatal day 21)

·         Two-month transmitter suitable for rats as young as postnatal day 7 and older

·         Two-month transmitter suitable for mice as young as postnatal day 12 and older

·         Six-month transmitter suitable for adult rats and mice

·         Transmitter mounts to the skull with cyanoacrylate

·         Reduces surgical time to tens of minutes

·         Teflon insulated platinum electrode wires are shipped at 10 cm and can be truncated to desired length; custom longer lengths available on request.

Transmitter specifications

Implant Weight 2 week – 0.5 g
2 month – 2.3 g
6 month – 4 g
Footprint 2 week – 4 mm x 6 mm
2 month – 7 mm x 9 mm
6 month – 7 mm x 12 mm
Volume 2 week – 0.192 cubic centimeter
2 month – 0.756 cubic centimeter
6 month – 1.344 cubic centimeter
Input Impedance 5 MΩ
Bandwidth 0.1 100 Hz per channel
Operating Temperature 30-45ºC
Implant Material Medical Grade Epoxy


Gain Options

This sensor ships with the default transmitter Gain setting; other Gain settings are available if requested before order is placed. AcqKnowledge software includes the scale settings for each transmitter gain option.


Typical Use

Gain

Settings

EEG, ECoG, LFP – default

2000x

±1.0 mV range, 2.0 mVpp in = 4 V out

Status-Epilepticus

800x

±2.5 mV range, 5.0 mVpp in = 4 V out

LFP, EEG in rat or mouse pups

4000x

 ±0.5 mV range, 1.0 mVpp in = 4 V out


Activate Sensors When You’re Ready To Use Them

Add the Sensor Activation & Test Utility (EPOCH-ACTI) to start the sensor battery when ready to use the transmitter.

  • The reusable activator is battery operated and can be used for any sensors that are ordered as deactivated.
  • Transmitters must be activated within 6 months of shipment to ensure the full 2-month or 6-month active battery-life.
  • Sensors will be shipped with the battery running unless requested as “deactivated” when ordered.
    • Request “deactivated sensors” in the comments section when ordering sensors or systems.
    • Deactivated sensors require that the user has an EPOCH-ACTI on-hand or includes one in the EPOCH order.
  • The activator utility can also output two independent channels of data (sinewaves) for testing purposes.